OBS500

Smart Turbidity Meter with Antifouling Features
OBS500

OBS500

Retired

Smart Turbidity Meter with Antifouling Features

No longer available to order

This product is retained for service, documentation, and compatibility information.

Replaced by

  • ClariVue 10

    ISO 7027-Compliant, Side-Scatter Turbidity Sensor

  • ClariVue 20

    ISO 7027-Compliant, Side-Scatter Turbidity Sensor

Model:
OBS500
Item Number:
27784
Warranty:
1 year

Overview

The OBS500 is a submersible turbidity probe with active antifouling capabilities for better measurements in biologically active water with both high and low turbidity. It outputs an SDI-12, digitally processed signal that many of our data loggers can measure. ClearSensor®  and OBS® are registered trademarks of Campbell Scientific.

Key Advantages

Accurate, Rugged
ClearSensor antifouling features, dual backscatter sensors

Technical Description

Design features of the OBS500 include the combination of a backscatter sensor (better at measuring higher turbidity) with a second sidescatter sensor (better at measuring lower turbidity). It has a shutter that is opened only during measurements, which reduces the time that algae or other organisms can cling to its optics.

To prevent biofouling and ensure better measurements, the OBS500 incorporates the ClearSensor Method (U.S. Patent No. 8,429,952). This method uses a shutter/wiper mechanism to protect and clean the optics. With the ClearSensor method, a chamber is also filled with a biocide that continuously leaches out over the optics while the probe shutter is in the closed position.

Campbell Scientific offers a disposable, plastic sleeve that can make cleanup a snap, as well as a copper sleeve that can provide additional protection, especially in sea water.

(OBS® is a registered trademark of Campbell Scientific.)

Benefits

  • Dual backscatter and sidescatter sensors used to measure turbidity
  • Patented ClearSensor antifouling method for better measurements in biologically active water
  • Shutter/wiper mechanism keeps lenses clean
  • Refillable biocide chamber prevents fouling
  • Disposable plastic sleeve facilitates cleanup
  • Optional copper sleeve for additional protection (especially for sea water) or disposable plastic sleeve facilitates easy cleanup

Specifications

Dual Probe
90° sidescatter and backscatter
Range
0 to 4000 NTU
Active and Passive Antifouling
Shutter, wiper, biocide, copper, optional removable sleeve
Concentration Accuracy
±2% of reading or 0.5 NTU

(whichever is greater)

Operating Temperature Range
0° to 40°C
Temperature Accuracy
±0.3°C
Emitter Wavelength
850 nm
Power Requirements
9.6 to 18 Vdc
Measurement Time
< 10 s
Maximum Submersion Depth
100 m

(330 ft)

Diameter
4.8 cm

(1.88 in.)

Maximum Cable Length
> 500 m

(1640 ft)

Length
27 cm

(10.63 in.)

Weight
0.59 kg

(1.3 lb)

Power Consumption

Quiescent
< 200 µA
Measurement
< 40 mA
Communication
< 40 mA
Maximum Peak Current
200 mA for 50 ms when shutter motor starts
Active Shutter Motor
< 120 mA

Outputs

SDI-12
Version 1.3, 1200 bps
RS-232
9600 bps
Analog
0 to 5 Vdc

Ships With

(2)

Related Resources

Product Life Events

Available for Order
July 28, 2011
Retired
June 21, 2018
Free Support Until
December 31, 2020
End of Life
December 31, 2020
Calibration Available Until
December 31, 2020
Repair Available Until
December 31, 2020