EC155
EC155
OrderableConfigurableClosed-Path CO2/H2O Gas Analyzer
Pricing
Configuration
Sign in for option pricing- Gas Analyzer Only No Sonic Case w/Basic Barometer w/EC155 Carrying Case-GH
- CSAT3A & Gas Analyzer w/Sonic Carrying Case w/Basic Barometer w/EC155 Carrying Case-SS
Select a configuration to add to your quote
- Model:
- EC155
- Item Number:
- 45305
- Warranty:
- 3 years or 17,500 hours
Overview
Campbell Scientific's EC155 closed-path analyzer incorporates vortex technology for reduced maintenance, an absolute pressure sensor in the sample cell for more accurate measurements, and a sample cell with improved corrosion protection. The EC155 can be combined with the CSAT3A sonic anemometer, as shown in the main image. The revised CSAT3A has a more aerodynamic and rigid design.
The EC155 is ordered as part of a CPEC300-series system (CPEC300, CPEC306, or CPEC310), which also includes the sample pump, data logger, optional valve module, and optional scrub module to provide a zero air source. The EC155 with anemometer simultaneously measures absolute carbon dioxide and water vapor mixing ratio, sample cell temperature and pressure, and three-dimensional wind speed and sonic air temperature.
Key Advantages
Use as part of closed-path eddy-covariance system
Technical Description
The EC155 has the following outputs:
- Ux (m/s)*
- Uy (m/s)*
- Uz (m/s)*
- Sonic Temperature (°C)*
- Sonic Diagnostic*
- CO2 Mixing Ratio (µmol/mol)
- H2O Mixing Ratio (mmol/mol)
- Gas Analyzer Diagnostic
- Cell Temperature (°C)
- Cell Pressure (kPa)
- CO2 Signal Strength
- H2O Signal Strength
- Differential Pressure (kPa)
- Source Temperature (°C)
*Requires a CSAT3A Sonic Anemometer Head.
Benefits
- Vortex Intake (U.S. Pat. No. 9,217,692) greatly reduces maintenance frequency compared to traditional in-line filters
- Heated inlet increases protection against condensation
- More accurate pressure measurements with the new sample cell absolute pressure sensor
- Fully integrated, detachable intake
- Improved corrosion protection with stainless-steel sample cell
- Improved sonic temperature from more rigid CSAT3 geometry
- Stream-lined, aerodynamic CSAT3A mounting
- Slim aerodynamic shape with minimal wind distortion
- Analyzer, sample cell, and sonic anemometer measurements have matched bandwidths and are synchronized by a common set of electronics
- Low power consumption; suitable for solar power applications
- Low noise
- Small sample cell for excellent frequency response
- Integrated zero/span connection for simplified field zero/span
- Field rugged
- Field serviceable
- Factory calibrated over wide range of CO<sub>2</sub>, H<sub>2</sub>O, pressure and temperature in all combinations encountered in practice
- Extensive set of diagnostic parameters
- Fully compatible with Campbell Scientific data loggers; field setup, configuration, and field zero and span can be accomplished directly from the data logger
- Rain: innovative signal processing and transducer wicks considerably improve performance of the anemometer during precipitation events
Specifications
- Operating Temperature Range
- -30° to +50°C
- Operating Pressure
- 70 to 106 kPa
- Input Voltage Range
- 10 to 16 Vdc
- Power
- 5 W
(steady state and power up) at 25°C
- Measurement Rate
- 60 Hz
- Output Bandwidth
- 5, 10, 12.5, or 20 Hz
(user-programmable)
- Output Options
- SDM, RS-485, USB, analog
(CO2 and H2O only)
- Auxiliary Inputs
- Air temperature and pressure
- EC100 Barometer Accuracy
- ±1.5 kPa (> 0°C), increasing linearly to ±.3.7 kPa at -30°C
(basic)
- EC100 Barometer Accuracy
- ±0.15 kPa (-30° to +50°C)
(enhanced)
- Sample Intake/Sonic Volume Separation
- 15.6 cm
(6.1 in.)
- Warranty
- 3 years or 17,500 hours of operation
(whichever comes first)
- Cable Length
- 3 m
(10 ft) from EC155/CSAT3A to EC100
- Weight
- 3.9 kg
(8.5 lb) for EC155 head and cables
- Weight
- 1.7 kg
(3.7 lb) for CSAT3A head and cables
- Weight
- 0.4 kg
(0.9 lb) for mounting hardware
- Weight
- 3.2 kg
(7 lb) for EC100 electronics
Gas Analyzer
- Sample Cell Thermistor Accuracy
- ± 0.15°C
(-30° to +50°C)
- Sample Cell Pressure Accuracy
- ± 1.5 kPa (> 0°C ), increasing linearly to ±3.7 kPa at -30°C
Gas Analyzer - CO<sub>2</sub> Performance
- Accuracy
- Assumes the following:
the gas analyzer was properly zero and spanned using the appropriate standards; CO2 span concentration was 400 ppm; H2O span dewpoint was at 12°C (16.7 ppt); zero/span temperature was 25°C; zero/span pressure was 84 kPa; subsequent measurements made at or near the span concentration; temperature is not more than ±6°C from the zero/span temperature; and ambient temperature is within the gas analyzer operating temperature range.
- Accuracy
- 1%
(Standard deviation of calibration residuals.)
- Precision RMS (maximum)
- 0.15 µmol/mol
Nominal conditions for precision verification test: 25°C, 86 kPa, 400 μmol/mol CO2, 12°C dewpoint, and 20 Hz bandwidth. - Calibrated Range
- 0 to 1,000 μmol/mol
(0 to 3,000 µmol/mol available upon request.)
- Zero Drift with Temperature (maximum)
- ±0.3 μmol/mol/°C
- Gain Drift with Temperature (maximum)
- ±0.1% of reading/°C
- Cross Sensitivity (maximum)
- ±1.1 x 10-4 mol CO2 /mol H2O
Gas Analyzer - H<sub>2</sub>O Performance
- Accuracy
- Assumes the following:
the gas analyzer was properly zero and spanned using the appropriate standards; CO2 span concentration was 400 ppm; H2O span dewpoint was at 12°C (16.7 ppt); zero/span temperature was 25°C; zero/span pressure was 84 kPa; subsequent measurements made at or near the span concentration; temperature is not more than ±6°C from the zero/span temperature; and ambient temperature is within the gas analyzer operating temperature range.
- Accuracy
- 2%
(Standard deviation of calibration residuals.)
- Precision RMS (maximum)
- 0.006 mmol/mol
Nominal conditions for precision verification test: 25°C, 86 kPa, 400 μmol/mol CO2, 12°C dewpoint, and 20 Hz bandwidth. - Calibrated Range
- 0 to 72 mmol/mol
(38°C dewpoint)
- Zero Drift with Temperature (maximum)
- ±0.05 mmol/mol/°C
- Gain Drift with Temperature (maximum)
- ±0.3% of reading/°C
- Cross Sensitivity (maximum)
- ±0.1 mol H2O/mol CO2
Sonic Anemometer - Accuracy
- -NOTE-
- The accuracy specification for the sonic anemometer is for wind speeds < 30 m s-1 and wind angles between ±170°.
- Offset Error
- < ±8.0 cm s-1
(for ux, uy)
- Offset Error
- < ±4.0 cm s-1
(for uz)
- Offset Error
- ±0.7° while horizontal wind at 1 m s-1
(for wind direction)
- Gain Error
- < ±2% of reading
(for wind vector within ±5° of horizontal)
- Gain Error
- < ±3% of reading
(for wind vector within ±10° of horizontal)
- Gain Error
- < ±6% of reading
(for wind vector within ±20° of horizontal)
- Measurement Precision RMS
- 1 mm s-1
(for ux, uy)
- Measurement Precision RMS
- 0.5 mm s-1
(for uz)
- Measurement Precision RMS
- 0.025°C
(for sonic temperature)
- Measurement Precision RMS
- 0.6°
(for wind direction)
Accessories
(10)CABLEPCBL
2-Conductor 16 AWG Power Cable
OrderableCABLE4CBL
4-Conductor 22 AWG Cable with Drain
OrderableCABLE2TP
22 AWG, 2-Twisted Pair Cable with Drain
OrderableCABLE3TP
24 AWG, 3-Twisted Pair Cable with Drain
Orderable44299
Replacement Sonic Head for CSAT3A
OrderableEasyFlux WebS
EasyFlux Web Software, Self-Hosted
RetiredEasyFlux WebH
EasyFlux Web Software, Campbell Scientific-Hosted
RetiredEasyFlux PC
Eddy-Covariance Post-Processing PC Software
OrderableEasyFlux DL
Eddy-Covariance Data Logger Program
OrderableFluoreSens 10
Sun-Induced Chlorophyll Fluorescence (SIF) System
Orderable
Need Something Not Listed?
(45305)Don't see what you need from among these options? Download the complete configuration list below.
View all product configurationsRelated Resources
Case Studies
Documents
- Technical Paper
Molecular Sieve CO2 and H2O Scrubbing Solution
September 2017
View Document - Brochure
EC155 CO2 and H2O Closed-Path Gas Analyzer
April 2026
View Document - Document
Using Molecular Sieve to Zero Infrared Gas Analyzers for Eddy Covariance or Atmospheric Profile Measurements
View Document - Brochure
Micrometeorology
August 2017
View Document - View all documents