ISFET pH Probe
ISFET pH Probe
RetiredNo longer available to order
This product is retained for service, documentation, and compatibility information.
Replaced by
- CS526-L
Digital ISFET pH Sensor
- Warranty:
- 6 months
Overview
Campbell Scientific’s CS525 pH probe provides reliable, accurate pH measurements that are fully temperature compensated. This probe makes general pH measurements in aqueous or semi-solid solutions, and can be submersed or inserted into tanks, pipelines, and open channels.
Key Advantages
Elimination of glass bulb allows longer time in field
Technical Description
The CS525 uses SENTRON’s high-tech, Ion Sensitive Field Effect Transistor (ISFET) semi-conductor as its pH-sensitive element, and includes a silver/silver chloride-potassium chloride reference system. The ISFET technology is the most powerful pH monitoring technology available today. This technology considerably reduces the number of acidic or alkaline errors in extreme pH conditions. It allows the CS525 to monitor pH in liquids containing high solids, aggressive chemicals, or biological materials that would clog or contaminate the junction of the traditional glass-bulb pH probes.
The CS525’s rugged design makes it suitable for just about any liquid pH-monitoring application, from laboratory to harsh field applications. The probe’s electronics are safely embedded in a durable PEEK body. Elimination of the glass-bulb removes the possibility of broken glass, making the CS525 more durable and safer to use.
| Note: The CS525-L ISFET pH Sensor is highly susceptible to ground looping and galvanic conditions. The best isolation technique to overcome these issues is to use a CWS900-series Wireless Sensor Interface. In many cases, the CS525 can also be isolated by NOT grounding the data logger to an earth ground. Please discuss your application with Campbell Scientific before purchasing the sensor. |
Product Notes
Benefits
- Compatible with the CWS900-series interfaces, allowing it to be used in a wireless sensor network
- Electronics are safely embedded in a durable resin body
- Elimination of the glass-bulb makes the probe more durable and safer to use
- Fully temperature-compensated
- Innovative ISFET pH-sensing element used as measurement electrode
- Designed and manufactured under stringent quality control conditions in an ISO 9001 environment, ensuring good measurements
- Each sensor individually tested
- Compatible with all contemporary data loggers (including the CR200X series)
Specifications
- CE Compliance
- CE Compliant
- pH Range
- 0 to 14
- Power Requirements
- 4.5 to 5.5 Vdc
- Current Consumption
- 2 mA
(nominal)
- Measurement Time
- 200 ms < 67% < 500 ms
- Accuracy
- ±0.1 pH
(with two-point calibration)
- Operating Temperature
- 0° to 70°C
- Allowed Water Pressure
- 0 to 30 kPa
(0 to 45 psi)
- Cable Type
- Two-twisted pair, 22 AWG cable with Santoprene jacket
- Sensor Material
- Polyaryletheretherketone (PEEK)
- Diameter
- 16 mm
(0.63 in.)
- Maximum Cable Length
- 300 m
(1000 ft)
- Length
- 190 mm
(7.48 in.)
- Weight
- 318 g (11.2 oz) with 3.05-m (10-ft) cable
Output
- -NOTE-
- Output range used depends on which wires are connected to the data logger.
- Range A
- 0 to 5000 mV+2700 mV at pH7 and 25°C with a sensitivity of 270 mV to 330 mV/pH at 25°C
- Range B
- ±400 mV with a sensitivity of -45 mV to -50 mV/pH at 25°C
Other Accessories
(2)7421
Split Mesh Cable Grip, 0.230 to 0.320
The 7421 Split Mesh Cable grip is recommended for use in wells to center the cable and to provide a method of suspending the cable, reducing cable stretch.
Orderable
VDIV2:1
2-to-1 Voltage Divider Terminal Input Module
The VDIV2:1 2-to-1 voltage divider is required if using a CR200-series, CR500, CR510, or CR10(X) datalogger to measure the probe.
Orderable
Similar Products
(1)Compatibility
Note: The following shows notable compatibility information. It is not a comprehensive list of all compatible or incompatible products.
Compatible Products
(11)
CR200X
Datalogger
RetiredNotesThe CR200X-series dataloggers require the VDIV2:1 2-to-1 Voltage Divider; contact Campbell Scientific about using this sensor with a retired data logger.

CR206X
Datalogger with 900 MHz Spread-Spectrum Radio
RetiredNotesThe CR200X-series dataloggers require the VDIV2:1 2-to-1 Voltage Divider; contact Campbell Scientific about using this sensor with a retired data logger.

CR211X
Datalogger with 922 MHz Spread-Spectrum Radio
End of lifeNotesThe CR200X-series dataloggers require the VDIV2:1 2-to-1 Voltage Divider; contact Campbell Scientific about using this sensor with a retired data logger.

CR216X
Datalogger with 2.4 GHz Spread-Spectrum Radio
RetiredNotesThe CR200X-series dataloggers require the VDIV2:1 2-to-1 Voltage Divider; contact Campbell Scientific about using this sensor with a retired data logger.

CR295X
Datalogger for GOES Satellite Communication
RetiredNotesThe CR200X-series dataloggers require the VDIV2:1 2-to-1 Voltage Divider; contact Campbell Scientific about using this sensor with a retired data logger.

CR800
Measurement and Control Datalogger
RetiredNotes—
CR850
Measurement and Control Datalogger
RetiredNotes—
CR1000
Measurement and Control Datalogger
RetiredNotes—
CR3000
Measurement and Control Datalogger
RetiredNotes—
CR5000
Measurement and Control Datalogger
RetiredNotes—
CR9000X
Measurement and Control Datalogger
RetiredNotes—
Related Resources
Product Life Events
- Available for Order
- May 20, 2009
- Retired
- November 1, 2011
- Free Support Until
- March 31, 2018
- End of Life
- March 31, 2018
- Calibration Available Until
- March 31, 2018
- Repair Available Until
- March 31, 2018