45223
45223
OrderableCS451 Pressure Transducer for OEMs
Pricing
Pricing temporarily unavailable.
Configuration
Sign in for option pricing- 0-2.9psig (20kPa) No Calibration Cert –SN w/Standard Nose Cone No OEM Label w/10m (33ft) per SensorCS451OEM-SA-2-NC-SN-NL-10m
- 0-7.25psig (50kPa) No Calibration Cert –SN w/Standard Nose Cone No OEM Label w/10m (33ft) per SensorCS451OEM-SA-7-NC-SN-NL-10m
- 0-14.5psig (100kPa) No Calibration Cert –SN w/Standard Nose Cone No OEM Label w/15m (50ft) per SensorCS451OEM-SA-14-NC-SN-NL-15m
- 0-29psig (200kPa) No Calibration Cert –SN w/Standard Nose Cone No OEM Label w/30m (100ft) per SensorCS451OEM-SA-29-NC-SN-NL-30m
- 0-72.5psig (500kPa) No Calibration Cert –SN w/Standard Nose Cone No OEM Label -60m w/60m (200ft) per SensorCS451OEM-SA-72-NC-SN-NL-10m
- 0-145psig (1000kPa) w/Calibration Cert w/NPT Fitting Company F OEM Label w/60m (200ft) per SensorCS451OEM-SA-145-CC-NN-FL-60m
Select a configuration to add to your quote
- Model:
- 45223
- Item Number:
- 45223
- Warranty:
- 1 year
Overview
This version of the CS451 pressure transducer is for OEMs. It does not have a label so that the OEM can attach their own branding.
Technical Description
The 45223 consists of a piezoresistive sensor and a temperature sensor housed in a 316L stainless-steel case. It has a rugged Hytrel cable that remains flexible, even under harsh environmental conditions. The cable incorporates a vent tube to compensate for atmospheric pressure fluctuations. The vent tube terminates inside a desiccant tube, which prevents water vapor from entering the inner cavity of the transducer.
The 45223 has several pressure range options and two accuracy options (see Ordering). The standard accuracy option provides ±0.1% FS TEB over the 0° to 60°C temperature range. The high accuracy option provides ±0.05% FS TEB over the 0° to 60°C temperature range and includes a calibration certificate. TEB is the combined errors due to nonlinearity, hysteresis, non-repeatability, and thermal effects over the compensated temperature range, per ISA S51.1. Please note that the high accuracy option is not available for the 0 to 2.9 psig range option.
Campbell Scientific offers the A150 Desiccated Case that allows the CS451 to be connected to a CWS900 Wireless Sensor Interface (for use in a wireless sensor network) or to be connected to a prewired enclosure (see Ordering Information).
Benefits
- Quality construction ensures product reliability
- Rugged stainless-steel case protects piezoresistive sensor
- Compatible with most Campbell Scientific dataloggers
- Fully temperature-compensated
- Low-power sleep state between measurements reduces power consumption
- Optional weighted nose cone facilitates submersion
- NPT fitting allows it to be used in closed-pipe applications
- Quick shipment after receipt of order (ARO)
- Compatible with the A150 Desiccated Case, which allows it to be connected to the CWS900 and used in a wireless sensor network
Specifications
- Measurement Time
- < 1.5 s
- Outputs
- SDI-12 (version 1.3) 1200 bps; RS-232 9600 bps
- Resolution
- 0.0035% FS
- Operating Temperature
- -10° to +80°C
WARNING: Sensor could be damaged if encased in frozen ice. - Compensated Temperature
- 0° to 60°C
WARNING: Sensor could be damaged if encased in frozen ice. - Temperature Accuracy
- ±0.2°C
- Overpressure
- 2 x pressure range
- Power Requirements
- 6 to 18 Vdc
- Cable Type
- Hytrel Jacket, five conductor, 26 AWG
- NPT Fitting
- 1/4-in. NPS
- Top Cone Material
- Delrin
- Body Material
- 316L stainless steel
- Element Material
- 316L stainless steel
- Distance
- 2.54 cm
(1 in.)
Distance from black line etched on housing to end of NPT fitting2.3 cm(0.9 in.)
Distance from black line etched on housing to end of standard nose cone9.9 cm(3.9 in.)
Distance from black line etched on housing to end of weighted nose cone - Diameter
- 21.34 mm
(0.84 in.)
- Length
- 213.36 mm
(6.875 in.)
- Cable Weight
- 0.0421 kg/m
(0.0283 lb/ft)
- Weight
- 0.17 kg
(0.37 lb) without cable
Power Consumption
- Quiescent
- < 50 µA
- Measurement/Communication
- 8 mA
(1 s measurement)
- Maximum
- 40 mA
Pressure Measurement Range
- Option -2
- ≤ 2.9 psig
(20 kPa)
- Option -7
- ≤ 7.25 psig
(50 kPa)
- Option -14
- ≤ 14.5 psig
(100 kPa)
- Option -29
- ≤ 29 psig
(200 kPa)
- Option -72
- ≤ 72.5 psig
(500 kPa)
- Option -145
- ≤ 145 psig
(1000 kPa)
Depth of Fresh Water Measurement Range
- Option -2
- ≤ 2.0 m
(6.7 ft)
- Option -7
- ≤ 5.1 m
(16.7 ft)
- Option -14
- ≤ 10.2 m
(33.4 ft)
- Option -29
- ≤ 20.4 m
(67 ft)
- Option -72
- ≤ 50.9 m
(167 ft)
- Option -145
- ≤ 102 m
(334.5 ft)
Accuracy
- Standard Accuracy Option
- ±0.1% full-scale-range TEB
Total Error Band (TEB) includes the combined errors due to nonlinearity, hysteresis, nonrepeatability, and thermal effects over the compensated temperature range, per ISA S51.1. - High Accuracy Option
- ±0.05% full-scale-range TEB
The high accuracy (±0.05% full-scale range) option is not available for the 0 to 2.9 psig range option.
Total Error Band (TEB) includes the combined errors due to nonlinearity, hysteresis, nonrepeatability, and thermal effects over the compensated temperature range, per ISA S51.1.
Maximum Cable Length
- SDI-12
- ~457 m
(1500 ft)
1 sensor connected to a single port60 m(200 ft)
10 sensors connected to a single port - RS-232
- 60 m
(200 ft)
Air Gap
- Standard & Weighted Nose Cone
- 0.653 cm
(0.257 in.)
- NPT Fitting
- 2.72 cm
(1.07 in.)
Accessories
(5)Need Something Not Listed?
(45223)Don't see what you need from among these options? Download the complete configuration list below.