How often does the T107 or the ET107 need to be calibrated?

Campbell Scientific recommends the following maintenance schedule:

  • Once a week: check the rain gage funnel, and clean as necessary.
  • Once a month: check the solar radiation sensor, and clean as necessary.
  • Every 3 months: clean the temperature/RH radiation shield.
  • Every 6 months to 1 year: replace the two desiccant packs (pn 4905).
  • Every 1 to 3 years: have Campbell Scientific replace the wind speed bearings and reed switch. Request a return material authorization (RMA) number by following the steps listed on our Repair and Calibration page.
  • Every 1 to 3 years: replace the relative humidity chip (pn 9598).
  • Once every 3 years: have Campbell Scientific calibrate the solar radiation sensor. Request a return material authorization (RMA) number by following the steps listed on our Repair and Calibration page.
  • Once every 5 years: have Campbell Scientific calibrate the rain gage sensor. Request a return material authorization (RMA) number by following the steps listed on our Repair and Calibration page.
  • Once every 5 years: have Campbell Scientific calibrate the data logger. Request a return material authorization (RMA) number by following the steps listed on our Repair and Calibration page.