39613

CR350 Calibration with Documentation
No image available

39613

Orderable

CR350 Calibration with Documentation

Model:
39613
Item Number:
39613

Overview

The 39613 provides full bench testing and calibration of the CR350 Measurement and Control Datalogger to meet original specifications. This process includes replacement of the internal lithium battery and desiccant pack, as well as an update of the operating system to the newest version. The CR350 is then tested over our standard temperature range (-40° to +70°C). A NIST-traceable calibration certificate is provided that shows both the "as received" and "as returned" measurements. After this service is completed, the CR350 is returned to you with all the settings restored to the default, unless otherwise specified. If the CR350 requires repair, those costs will be added to the invoice. A Return Material Authorization (RMA) from Campbell Scientific is required for the 39613 service.