v1.5July 2, 2025
Download unavailable
  • Improved reliability of SDI-12 when communicating with multiple sensors on cables longer than the length specified in the SDI-12 standard (tested at 300m).
  • Fixed issue that causes some sensors to fail to respond over SDI-12 at temperatures below -30degC.
  • Improved resolution of the mmHg result returned in the M3 command.
  • Fixed invalid CRC for RC! command when the sensor address is not 0.
v1.4April 17, 2023
Download unavailable
  • Fixed an issue where memory corruption could occur during a brownout event.
  • Implemented support for the usage of both line feed and/or carriage return characters during RS-232 communication.
v1.3May 5, 2021
Download unavailable
  • Initial release of the BaroVUE10 OS.