Resource Center

CS120 Firmware

Software applications, data logger operating systems, firmware, and drivers for Campbell Scientific products

CS120 Firmware

v6

Includes the current CS120 firmware and the CS120 support files for the Device Configuration Utility.

Revision History

v6August 10, 2012
Download unavailable
  • Support for internal power supply monitoring and associated alarm reporting added. (Only active with the latest hardware).
  • The internal dew heaters now do not turn on for 10 seconds after the sensor is powered up to help control initial inrush currents on power-up.
  • The dew heaters do not turn off now when the hood heaters turn ON at lower temperatures.
  • Fixed a bug where POLL requests would not be responded to occasionally.
  • Fixed a bug where the system status in the menu system would not refresh correctly.
  • Various performance improvements have been made to the serial communications system to improve response times.