CRS456
CRS456
OrderableTitanium Stand-Alone Pressure Transducer
Pricing
Configuration
Sign in for option pricing- 0-145psig (1000kPa) No Calibration Cert –SN w/Standard Nose ConeCRS456-SA-145-NC-SNPart Number: 45301-SA-145-NC-SN
- 0-145psig (1000kPa) w/Calibration Cert –SN w/Standard Nose ConeCRS456-SA-145-CC-SNPart Number: 45301-SA-145-CC-SN
- 0-14.5psig (100kPa) No Calibration Cert –SN w/Standard Nose ConeCRS456-SA-14-NC-SNPart Number: 45301-SA-14-NC-SN
- 0-14.5psig (100kPa) w/Calibration Cert –SN w/Standard Nose ConeCRS456-SA-14-CC-SNPart Number: 45301-SA-14-CC-SN
- 0-29psig (200kPa) No Calibration Cert –SN w/Standard Nose ConeCRS456-SA-29-NC-SNPart Number: 45301-SA-29-NC-SN
- 0-29psig (200kPa) w/Calibration Cert –SN w/Standard Nose ConeCRS456-SA-29-CC-SNPart Number: 45301-SA-29-CC-SN
- 0-72.5psig (500kPa) No Calibration Cert –SN w/Standard Nose ConeCRS456-SA-72-NC-SNPart Number: 45301-SA-72-NC-SN
- 0-72.5psig (500kPa) w/Calibration Cert –SN w/Standard Nose ConeCRS456-SA-72-CC-SNPart Number: 45301-SA-72-CC-SN
- 0-7.25psig (50kPa) No Calibration Cert –SN w/Standard Nose ConeCRS456-SA-7-NC-SNPart Number: 45301-SA-7-NC-SN
- 0-7.25psig (50kPa) w/Calibration Cert –SN w/Standard Nose ConeCRS456-SA-7-CC-SNPart Number: 45301-SA-7-CC-SN
Select a configuration to add to your quote
- Model:
- CRS456
- Item Number:
- 45301
- Warranty:
- 1 year
Overview
The CRS456 consists of a water-level and water-temperature sensor that has its own time clock and memory to store the collected data, in a compact titanium case. This frees users to place the sensor in remote sites and let it collect data for long periods. The sensor can then be retrieved, connected to a PC via a micro-USB port, and the data transferred to the PC. HydroSci software is included and elegantly supports test setup, data retrieval, and data display. Long battery life and rugged construction mean you can trust the CRS456 to collect important data. Low cost and ease of use make it a good choice in a variety of applications. The CRS451 is the same as this, but with a stainless-steel case.
Key Advantages
Easy to use software, reliable hardware
Technical Description
The CRS456 has several pressure range options. The CRS456 is unvented; therefore it is monitoring sealed gage pressure. For maximum accuracy, the data should be corrected for barometric pressure.
HydroSci software is available for download. This software simplifies the process of configuring the CRS456. Users can configure the CRS456 to monitor surface water, ground water, or a standard pump test.
HydroSci software will also compensate the measurements for barometric pressure using an additional CS456 or a CS451 as a reference sensor, or using a barometric pressure sensor from a nearby weather station.
HydroSci software will display the data in tabular or graphical formats.
Benefits
- Sensors and data-collection features in one instrument case
- Rugged titanium case protects piezoresistive sensor allowing it to be used in saltwater and other harsh environments
- Quality construction ensures product reliability
- Fully temperature-compensated
- Fast scan rate
- Large data-storage capacity
- Long battery life
- Easy-to-use software
Specifications
- Venting
- Unvented
- Measurement Time
- < 1.0 s
- Output Options
- micro USB
- Internal Data Collection Memory
- 4 MB
- Logging/Scanning Modes
- Standard, Delta, Wave, Logarithmic
- Resolution
- 0.0035% FS
- Dry Storage Temperature Range
- -10° to +80°C
WARNING: Sensor could be damaged if encased in frozen ice. - Operating Temperature Range
- 0° to 60°C
WARNING: Sensor could be damaged if encased in frozen ice. - Overpressure
- 2 x pressure range
- Power Requirements
- Internal user-replaceable lithium battery
- Battery Life
- 5+ years
(when logging interval is once per hour)
- Body Material
- Titanium
- Distance
- 2.54 cm
(1 in.)
Distance from black line etched on housing to end of NPT fitting2.3 cm(0.9 in.)
Distance from black line etched on housing to end of standard nose cone9.9 cm(3.9 in.)
Distance from black line etched on housing to end of weighted nose cone - Water-Level Accuracy
- ±0.1% full-scale-range TEB
Includes the combined errors due to nonlinearity, hysteresis, nonrepeatability, and thermal effects over the compensated temperature range, per ISA S51.1. - Temperature Accuracy
- ±0.2°C
- Ingress Protection
- Exceeds IP68
- Diameter
- 2.22 cm
(0.875 in.)
- Length
- 22.23 cm
(8.75 in.)
- Weight
- 230 g
(0.51 lb)
Power Consumption
- Quiescent
- < 80 µA
- Measurement/Communication
- 4 mA
(1 s measurement)
Measurement Ranges at Fresh Water Depths
- 0 to 5.1 m (16.7 ft)
- 0 to 7.25 psi0 to 50 kPa
- 0 to 10.2 m (33.4 ft)
- 0 to 14.5 psi0 to 100 kPa
- 0 to 20.4 m (67 ft)
- 0 to 29 psi0 to 200 kPa
- 0 to 50.9 m (167 ft)
- 0 to 72.5 psi0 to 500 kPa
- 0 to 102 m (334.5 ft)
- 0 to 145 psi0 to 1000 kPa
Common Accessories
(3)45006
Suspension Cable Kit for CRS451 or CRS456
The 45006 is a suspension cable kit used to deploy the CRS456.
Orderable31648
Heyco Cable Grip Liquid Tight
The 31648 is a clamp with 1 in. NPT threads that allows the CRS456 to be mated with a 1 in. PVC pipe for installations purposes. A 1 in. female adapter socket with FPT threads is needed between the 31648 and 1 in. PVC pipe. This adapter socket is typically available at stores that sell PVC pipe.
Orderable39642
SDI-12 DIN Rail Terminal Bus Kit without Mounting Bracket
The 39642 makes it easier for you to connect wires from multiple SDI-12 sensors into one set of terminals on a data logger. The 39642 uses six DIN-rail mountable (pn 27479), 6-wire terminal blocks. The 39642 is a jumper that electrically connects two terminal blocks together at a time (pn 32918). This setup allows up to 11 SDI-12 sensors to be connected to one set of terminals on a data logger.
Orderable
Replacement Parts
(2)29421
Lithium 3.6V AA Battery Pack for CRS451/V or CRS456/V
The 29421 is a replacement lithium battery pack for the CRS456.
Orderable27555
USB 2.0 Cable, Type A Pin (Male) to Micro B Pin (Male), 6/6.5 ft
The 27555 is a USB Type A to Micro B, pin (male)-to-pin (male) cable, 6 to 6.5 ft long. It connects devices using the Micro B interface to a USB port on a computer.
Orderable
Need Something Not Listed?
(45301)Don't see what you need from among these options? Download the complete configuration list below.
View all product configurationsSimilar Products
(6)CRS451
Stainless-Steel Stand-Alone Pressure Transducer
OrderableCS456
Titanium Pressure Transducer
OrderableCS451
Stainless-Steel Pressure Transducer
OrderableCRS451V
Stainless-Steel Vented Stand-Alone Pressure Transducer
OrderableCRS456V
Titanium Vented Stand-Alone Pressure Transducer
OrderableLevelVue B10
Water-Level Continuous Flow Bubbler with Integrated Screen
Orderable
Compatibility
Communications
Requires a micro USB cable and HydroSci software running on a computer with Windows XP or later. HydroSci software is available for download.
Related Resources
Documents
Downloads
- HydroSci Chinese Languagev1.2Login required
January 30, 2017
Software ApplicationsAdds Chinese language support for version 1.2 of HydroSci. Note: Language support for the above listed version only.
- HydroSciv1.2.1Login required
October 9, 2017
Software ApplicationsHydroSci is used to configure and extract data from the CRS45X Recording Submersible Pressure Transducer and the OBS-3A Turbidity and Temperature Monitoring System. Watch the Video Tutorial: Using HydroSci software with water-level recording sensors.