General Support
User Forum
Visit the Campbell Scientific Forum to submit your question to users around the world.
Ask an Expert
Ask a Question using our simple form and we'll pass it along to one of our experts.
Call Us
Our Application Engineers are ready to answer your questions. Our support numbers are:
Technical Support
US Sales
International Sales
Order Entry
Accounts Receivable
Repairs
|
435.227.9100
435.227.9120
435.227.9140
435.227.9090
435.227.9092
435.227.9105
|
Downloads
There are currently no Downloads for the 229-L.
FAQ
Q. What is the stated accuracy for the thermocouple (temperature measurement) in a 229 Water Matric Potential Sensor?
If all the errors were summed in a worst case scenario, the thermocouple accuracy would be ±0.5° C. For an extended discussion about thermocouple measurements, please refer to the "Thermocouple Measurements" section in the datalogger manual.
..more
Q. I have several 229-L probes installed at a site. They've been in operation for about 5 years. In the last 3 months, two of the sensors have begun returning very low positive and negative values for dT. The recorded temperature at the same depth seems OK, however. Is this an indication the heater is malfunctioning? Is there anything I can do to fix the probe without digging it back up? Is there possibly something else wrong that might be easier to fix? There have been no changes to the data logge
Very low values for dT, especially negative values, do indicate that
the 229 sensor is not heating during the measurement time. Because the
sensors were working fine but suddenly stopped working without a change
to the datalogger program, this indicates a physical problem. The most
likely causes are damage to the heater wires, one or more loose wires,
or a failure of the constant current excitation module. The first check
to make is the heater wire itself. With an ohm-meter, measure the
resistance between the black and green wires on the 229 sensor. The
resistance should read about 33-36 ohm. An off-scale or infinite reading
indicates a break in the heater wire. A low reading could indicate a
short in the heater wire. In both of those cases, the fix would be to
dig up the sensor wire and examine it for damage. The next check is to
make sure that there are good electrical connections where the black and
green sensor wires connect to the constant current interface or to a
multiplexer, where the multiplexer common wires connect to the constant
current interface, and where the constant current interface connects to
the datalogger 12-V, G, and control port. Each of those wires should
have about a quarter inch of bare copper securely connected to its
terminal. Finally, the CE4 or CE8 constant-current module can be checked
with a multimeter to make sure it is putting out 50 mA. Measure the
voltage between the 12-V and ground terminal screws to make sure that
the constant current module is receiving power from the datalogger.
Next, temporarily move the wire connecting the datalogger control port
to the CTRL channel to a 5-V channel on the datalogger to force the
module to be on all the time so you can measure the current output. Set
the multimeter to measure milliamps, and measure the current. It should
be 50 mA ± 1 mA. If the current output is outside that range, the
constant current module needs to be repaired or replaced.
..more
Q. How do I actually get my reading off of the device?
See Section 5 of the 229 manual
(campbellsci.com/documents/manuals/229.pdf) for details and sample programs.
..more
Q. Can you explain, with as much detail as possible, how we calibrate each sensor individually?
See section 6 of the 229 manual
(campbellsci.com/documents/manuals/229.pdf) for details.
..more
Q. Do you offer calibration of the 229-L heat dissipation probe for soil water matric potential?
At this time Campbell Scientific does not offer a calibration service for the 229 probe. There are companies that offer this service. Contact Campbell Scientific for details.
..more
Q. I am considering using the 229-L matric potential sensor with a CR-1000 datalogger. Can I program the system to read equilibrium soil temperature prior to excitation for matric potential measurement? That way I measure soil temperature and matric potential with on sensor.
Yes, we recommend that you measure and record the temperature of the sensor prior to turning on the heater and that reading can be used as a soiltemperature reading. As long as matric potential measurements are not mademore frequently than once every 15 minutes, there will be no significantheat buildup in the sensor and the sensor temperature will be the same asthe soil temperature.
..more
Q. I gather that the heater resistance should be around 33 ohms. Do resistance values of 37-48 ohms indicate a problem with the sensor, or is that within the normal range?
The heater wire itself should read around 33 ohms resistance, but the green and black wires connected to the heater wire will also add resistance to your reading. The longer the sensor cable, the greater your resistance reading will be, so readings of 37-48 ohms are within the normal range. A resistance reading of infinity or of less than 33 ohms would be caused by a break or short in the sensor cable and would be cause for concern.
..more
Q. What is the the 229-L heating element made of? In a acidic environment (copper mine tailings), would you expect it to fail due to corrosion in five years?
The heating element is made of Evanohm wire. The element is encased in epoxy inside a stainless steel hypodermic needle and will not be exposed to the corrosive environment, so that part of the sensor should not fail in five years. However the sensor cable and the ceramic matrix might suffer damage after five years in a corrosive environment.
..more
|